I 12 Fluid Responsiveness

12_tocConfig.jpg

Chapter Authors & Contributors


12.1 Introduction

Administration of an intravenous fluid challenge is a common medical intervention in the hypotensive or hypovolemic patient. Ideally, a fluid challenge increases stroke volume (SV), and thereby increases cardiac output. However, in some circumstances, a fluid challenge may be ineffective or harmful.[1][2][3] A patient receiving fluid for septic shock may not be fluid responsive after receiving multiple fluid challenges. A patient in cardiogenic shock with pulmonary edema may suffer additional harm by unnecessary administration of intravenous fluid. It is important that the clinician be able to properly assess a patient’s likelihood to respond to a fluid challenge. Newer methods incorporating ultrasound into the bedside evaluation may offer a superior assessment than conventional methods.

12.2 CVP and PAOP

Highlights

CVP is an inadequate assessment of fluid responsiveness.

Traditional methods of assessment of fluid response have relied on measuring cardiac filling pressures as a way to gauge preload. Central venous pressure (CVP) is a surrogate for right ventricular (RV) end-diastolic pressure, and pulmonary artery occlusion pressure (PAOP) is a surrogate for left ventricular (LV) end-diastolic pressure. Maestrini’s Law of the Heart (commonly called the Frank-Starling law) demonstrates that the relationship between ventricular end-diastolic pressure and ventricular end-diastolic volume (cardiac preload) is not linear (#Figure 12.1).[4][5] At low filling pressures, one may get large increases in SV for an incremental increase in filling pressure. At high pressures, one may get a minimal increase (or occasionally a decrease) in SV for the same increase in filling pressure. It is for this reason that CVP and pulmonary artery filling pressures are not accurate predictors of fluid responsiveness, despite their widespread clinical use.[6][7][8][9]

Figure 12.1 Frank Starling Curve

12_FigFrankStarling.jpg400

12.3 Assessing Fluid Responsiveness with U/S

Highlights

Advantages of ultrasound over traditional methods of assessing fluid response:

  • Visualization of ventricular chamber
  • Ability to quantify parameters that vary dynamically

During respiration, the changes in intrathoracic pressure affect preload.

Ultrasound offers several advantages over traditional methods of assessing fluid response. The ability to qualitatively visualize the ventricular chamber is sometimes more informative of cardiac preload than a measurement of its pressure. Perhaps more important is the ability of ultrasound to quantify parameters that vary dynamically with a perturbation of the cardiac filling pressures. The ability to measure dynamic parameters may inform the clinician whether the patient is situated on the steep part of the Frank-Starling curve or the plateau (#Figure 12.2).

Figure 12.2 Frank Starling Curve with Dynamic Parameters

12_FigFSCurveDynamicParams.jpg400

During respiration, the changes in intrathoracic pressure affect preload. In a passively mechanically ventilated patient who is fluid responsive, inspiration increases intrathoracic pressures, which reduces RV preload and SV. After two or three cardiac cycles, this decreased RV stroke volume leads to a decrease in LV preload and stroke volume, which are often observed during expiration. In the same patient, inspiration will lead to increased LV preload and SV due to increased pulmonary venous return of blood (#Figure 12.3). In a patient who is passively mechanically ventilated and fluid responsive, respiration will induce cyclical changes in SV. In a fluid responsive patient who is not receiving any positive pressure ventilation, inspiration will decrease intrathoracic pressures. SV will still vary with respiration, but will be reversed when compared to the changes noted in the passively mechanically ventilated patient.

Caution must be applied when evaluating the use of parameters that vary with respiration in the patient who is spontaneously breathing. The tidal volumes and deflections of intrathoracic pressures are inconsistent in the spontaneously breathing patient. Patients who are receiving an assisted mode of ventilation or non-invasive positive pressure ventilation may have more complex hemodynamic effects than either unassisted breathing or passive ventilation.[10][11]

Figure 12.3 Respiratory Variation of Stroke Volume

12_FigRSVariation.jpg400

12.4 Inferior Vena Cava

Highlights

  • Ultrasound evaluation of the IVC may inform the clinician of the CVP

  • However, CVP is not terribly useful to predict fluid responsiveness

  • In mechanically ventilated patients, a vena cava that distends >18% with passive respiration predicts fluid responsiveness

The inferior vena cava (IVC) is a highly compliant blood vessel that empties into the right atrium (RA). Ultrasound evaluation of the IVC may inform the clinician of the CVP[12] (#Table 12.1, #Movie 12.1, and #Movie 12.2).

Table 12.1 IVC and CVP

12_TableIVC&CVP.jpg

Movie 12.1 IVC Collapsing
Movie 12.2 Plethoric IVC

The utility of an ultrasonic estimation of CVP is limited, as CVP is not a terribly useful parameter to predict fluid responsiveness. The IVC runs longitudinally along the spine. The appropriate method to identify the IVC uses either a cardiac or abdominal 2-D ultrasound in the subcostal window. The measurement should be made perpendicular to the long axis of the IVC at end-expiration, just cephalad to the junction of the hepatic veins that lie approximately 5-30 mm proximal to the ostium of the RA (#Image 12.1). It is important to not mistake lateral translation of the IVC for collapse (#Movie 12.3). Evaluation of respiratory collapse is performed in the non-intubated patient by requesting that the patient perform a sniff maneuver. The degree of IVC collapse is not a reliable estimate of CVP in a patient receiving mechanical ventilation, but it is reasonable in these patients to expect that CVP will be <10 mm Hg if the IVC diameter is < 12mm.[13]

Image 12.1 Appropriate Location for IVC Measurement

12_ScanIVCMeasurementLocation.jpg400

Movie 12.3 Lateral Movement of IVC in Longitudinal Plane

Although the utility of an estimated CVP is limited, ultrasound can be very informative in predicting fluid response. Barbier and colleagues demonstrated that in mechanically ventilated patients, a vena cava that distends > 18% in diameter (#Figure 12.4) with passive respiration is likely to respond to a fluid challenge.[14] In a similar study comprised solely of septic patients receiving mechanical ventilation, Feissel and colleagues demonstrated that a passively ventilated patient with an IVC variability > 12% (#Figure 12.5) was likely to respond to fluid.[^15] The IVC may not be as informative in patients with abdominal compartment syndrome, pregnancy, or high airway pressures.

Equation 12.4 Vena Cava Distensibility
Vena Cava Distensibility=Dmax−DminDmin
Equation 12.5 Vena Cava Variability
Vena Cava Variability=Dmax−Dmin(Dmax+Dmin)/2

See the One Minute Ultrasound for the IVC measurement with respiratory variation (#Movie 12.4).

Movie 12.4 One Minute Ultrasound IVC Demonstration

12.5 Internal Jugular Vein

Highlights

There is growing interest in ultrasound of the IJ for fluid responsiveness; however, at the time of this publication, there is not sufficient evidence to support it.

There is growing interest in using ultrasonography of the internal jugular (IJ) vein to assess fluid responsiveness. Historically, clinical assessment of a patient’s volume status included an assessment of jugular venous pulsation, a measurement of marginal accuracy and reproducibility.[^16] One might surmise that ultrasound evidence of a collapsing or plethoric IJ vein would be as informative as the vena cava. Some clinicians in practice are using ultrasound to predict fluid responsiveness by respiratory collapse of the internal jugular. At the time of writing this text, there is insufficient evidence to support this practice.

12.6 Diastolic Function

Highlights

Diastolic function is helpful in determining the risk of a fluid challenge, but does not adequately assess fluid responsiveness

Apical 4-chamber view is used to obtain the mitral inflow velocity

The mitral annulus tissue dopller is also obtained in the apical 4-chamber view

Assessment of left ventricular diastolic function can be performed using pulse wave Doppler of the mitral inflow and the tissue Doppler of the mitral annulus (#Figure 12.6).[^17]

While not every patient’s diastology is easily categorized, one can use the ratio of mitral inflow velocity (E) to mitral annulus velocity (e’) during early diastole to estimate left atrial pressures.[^18]

Figure 12.6 Doppler of Mitral Valve Inflow and Tissue Doppler Imaging Patterns

12_FigPWD_MVI_TDITracings.jpg400

The mitral inflow velocity is obtained by transthoracic echocardiography in the apical 4-chamber (A4C) view or by transesophageal echocardiography in the midesophageal 4-chamber view (#Gallery 12.1).

The pulse wave Doppler, with a 1-3mm sample volume, is placed between the mitral leaflet tips during diastole.[^17] The mitral annulus tissue Doppler is obtained by transthoracic echocardiography in the A4C view, or by transesophageal echocardiography in the midesophageal 4-chamber view (#Gallery 12.1).

The tissue Doppler sample volume is positioned at or within 1 cm of the septal mitral annulus and adjusted to cover the longitudinal excursion of the mitral annulus (usually 5-10 mm).[^17] The E/e’ ratio offers a reasonable estimate of left atrial pressure. An E/e’ ratio < 8 is likely to correlate to a normal left atrial pressure. Conversely, an E/e’ ratio > 15 is likely to correlate to an elevated left atrial pressure.[^18] See the One Minute Ultrasound for diastolic function (#Movie 12.5).

Movie 12.5 One Minute Ultrasound Diastolic Function
[[12_FigDiastology1.jpg]] | The mitral inflow velocity is obtained by transthoracic echocardiography in the apical 4-chamber (A4C) view, or by transesophageal echocardiography in the midesophageal 4-chamber view.
[[12_FigDiastology2.jpg]] | The mitral inflow velocity is obtained by transthoracic echocardiography in the apical 4-chamber (A4C) view, or by transesophageal echocardiography in the midesophageal 4-chamber view.
[[12_FigDiastology3.jpg]] | The mitral annulus tissue Doppler is obtained by transthoracic echocardiography in the A4C view, or by transesophageal echocardiography in the midesophageal 4-chamber view.
[[12_FigDiastology4.jpg]] | The mitral annulus tissue Doppler is obtained by transthoracic echocardiography in the A4C view, or by transesophageal echocardiography in the midesophageal 4-chamber view.

Although an assessment of diastolic function and an estimate of left atrial pressure may not necessarily inform the clinician of the likelihood of the patient to respond to fluid, these data are useful when weighing the potential harms of administering a fluid challenge. A patient with an elevated left atrial pressure is at increased risk for pulmonary edema, a finding that can also be assessed with ultrasound (#Movie 12.6 and #Movie 12.7).[^19]

Movie 12.6 Lung Ultrasound Showing Pulmonary Edema
Movie 12.7 Normal Lung Ultrasound

12.7 Stroke Volume, Aortic Blood Velocity Variation

Highlights

  • Stroke volume can be measured using pulse wave Doppler of the LVOT in a parasternal long-axis view.
  • Stroke volume variation and aortic flow velocity variation predict fluid responsiveness in passively breathing, mechanically ventilated patients.

A clinician can determine left ventricular stroke volume using pulse wave Doppler of the left ventricular outflow tract (LVOT). The LVOT diameter is measured from a parasternal long-axis view (transthoracic) or midesophageal long-axis view (transesophageal) (#Figure 12.7, #Movie 12.8 and #Image 12.2).

Figure 12.7 Parasternal Long-Axis View

12_FigPSLView.jpg400

Movie 12.8 Parasternal Long-Axis View
Image 12.2 PSLA with LVOT Diameter

12_ScanPSLLVOT.jpg400

The velocity-time integral (VTI) of LVOT blood velocity is obtained using pulse wave Doppler from a transthoracic apical 5-chamber or apical long-axis view (#Image 12.3) or from a deep transgastric five-chamber view (#Movie 12.9 and #Image 12.4).

Image 12.3 LVOT VTI from Apical 5-Chamber

12_ScanA5CVTI.jpg400

Movie 12.9 LVOT VTI from Deep Transgastric
Image 12.4 LVOT VTI from Deep Transgastric

!500

Accurate assessment of SV requires that the pulse wave is placed at the same location where the LVOT diameter is measured. The angle of insonation can greatly affect the velocity of blood flow, which can alter the measured stroke volume (#Image 12.5 and #Figure 12.8). If the angle of insonation is < 15° from the direction of flow, the measured velocity is > 97% of the true velocity.

Image 12.5 Angle of Insonation

12_ScanAngleofInsonation.jpg400

Figure 12.8 Diagram of the Angle of Insonation, cosθ

12_FigErrorWAngleInsonation.jpg400

The clinician can use respiratory variation in SV to assess a patient’s likelihood to respond to fluid (#Image 12.6, #Image 12.7 and #Figure 12.9). As the LVOT diameter does not vary with respiration, stroke volume variation is essentially the variation in VTI. As stated above, the angle of insonation will affect the measurement of velocity. However, even if the clinician insonates with an angle >15° from the true direction of flow, it will scale down the measurement equally. Hence, a measurement of velocity that is off by 45° will be consistently decreased by 30% throughout the entire respiratory cycle. The stroke volume will not be accurate, but the SV variation will remain accurate.

Measuring stroke volume in real-time requires calculation of the velocity-time integral, which requires a software package capable of performing these calculations. In some instances, longitudinal movement of the heart may result in difficulty with appropriate placement performing these calculations. In some instances, longitudinal movement of the heart may result in difficulty with appropriate placement of the sample volume of the pulse wave Doppler. Pulse wave Doppler is also subject to aliasing, which may make measurement of the velocity-time integral difficult or impossible (#Image 12.8).

Image 12.6 LVOT VTI with Respiratory Variation

12_ScanLVOTRSVariation.jpg500

Image 12.7 LVOT without Respiratory Variation

12_ScanLVOTNoRSVariation.jpg 500

Equation 12.9 Stroke Volume Variability
Stroke Volume Variability=VTImax−VTImin(VTImax+VTImin)/2
Image 12.8 Aliased LVOT VTI

Perhaps a more useful surrogate measurement is the variation of maximum aortic blood velocity, using continuous wave Doppler. In most patients without significant aortic stenosis, the peak aortic blood velocity will be proportional to the stroke volume (#Images 12.9, #Image 12.10, and #Figure 12.10). Measurement of aortic blood velocity avoids the problems of aliasing and longitudinal movement of the heart. Another advantage is that a clinician may easily estimate peak velocity variation at bedside without use of a software package.

Image 12.9 Ao Velocity - Respiratory Variation

12_ScanPWDAoVRSVariation.jpg500

Image 12.10 Ao Velocity - No Respiratory Variation

12_ScanPWDAoVNoRSVariation.jpg500

Equation 12.10 Aortic Velocity Variability
Aortic Velocity Variability=AoVmax−AoVmin(AoVmax+AoVmin)/2

Ultrasound assessment of stroke volume variation, or its surrogate measurement of aortic velocity variation, appears to be informative of fluid responsiveness for select populations. In passively ventilated patients with septic shock, an aortic blood velocity variation threshold > 12% obtained by transthoracic echocardiography indicated a strong likelihood for fluid responsiveness.[^20] A similar study was performed by Monnet and colleagues, using transesophageal assessment of aortic velocity variation. In that study, passively ventilated ICU patients were likely to respond to fluid if their aortic velocity variation was > 18%.[^21]

The use of aortic velocity variation or echocardiographic stroke volume variation is not generalizable to all patients. These indices are not useful in a patient with an irregular ventricular rhythm, such as atrial fibrillation or frequent premature ventricular contractions. In such patients, SV will vary depending on diastolic filling time, and will not be informative of fluid responsiveness. For clinicians measuring Doppler velocities with transthoracic echocardiography, care must be taken to ensure that the angle of the probe remains the same throughout the respiratory cycle, as alteration in the angle of insonation may alter the measured velocity. In the same vein, a patient with significant cardiac translation (e.g., a patient with electrical alternans in the setting of a large pericardial effusion) may appear to have large variations in aortic blood velocity, when in reality the only thing that varies is the angle of insonation. Perhaps the most important limitation to the generalizability of ultrasonic assessment of stroke volume variation or aortic velocity variation is that the studies in this area are limited to passively breathing, mechanically ventilated patients. Thus, in a patient who is breathing spontaneously, these indices have not been validated. One study, by Skulec and colleagues, has looked at these indices in spontaneously breathing healthy volunteers who were synchronizing their breathing to a metronome to decrease breath-to-breath variability. In these patients, a stroke volume variation of > 17% or an aortic velocity variation >14% predicted fluid response.[^22]

12.8 Ventricular Size

Highlights

  • Hyperdynamic hypovolemia is usually associated with preload sensitivity
  • The literature surrounding the usefulness of visual estiation of the LV for fluid responsiveness in conflicting.

Ultrasound can easily obtain images of the heart, whether by transthoracic or transesophageal echocardiography. A clinician can identify hyperdynamic hypovolemia on echocardiography, a finding that is usually associated with preload sensitivity (#Movie 12.10 and #Movie 12.11). While it may at first seem that a visual estimate of the LV would be the best method of assessing fluid responsiveness, the literature surrounding this are conflicting. Reuter and colleagues demonstrated that left ventricular end-diastolic area was able to predict fluid responsiveness.[^23] Contrarily, multiple studies have demonstrated that echocardiographic measurement of the LV area or volume at end-diastole is not useful to predict fluid response.[^20][^24][^25][^26]

Movie 12.10 Normal Parasternal Short
##### Movie 12.11 Hyperdynamic Hypovolemic Parasternal Short

12.9 Passive Leg Raise

Highlights

  • CO and aortic velocity with passive leg raise is the “only maneuver shown to be accurate in predicting fluid responsiveness in spontaneously breathing patients.
  • Passive leg raise is effectively a reversible fluid bolus.
  • Increase in stroke volume >12.5% correlates with fluid responsiveness.

The passive leg raise is a maneuver performed at bedside, which will result in immediate distribution of approximately 300 mL of whole blood into the central venous system. The advantage of this maneuver is that it is effectively a reversible fluid challenge. It is done by assessing cardiac output or SV with the head of the bed elevated to 45°. The head of the bed is immediately lowered to 0° and the legs are elevated 45° for 2 minutes (#Figure 12.11 and #Figure 12.12).

Figure 12.11 Baseline Measurements

12_FigPLRBaseline.jpg400

Figure 12.12 Leg Raise Measurements

12_FigPLRElevated.jpg400

The cardiac output or stroke volume is then measured again. A patient who has a significant increase in either of these measurements is likely to respond to a fluid challenge. A key advantage of the passive leg raise is its applicability in the spontaneously breathing patient. Lamia and colleagues used the passive leg raise in spontaneously breathing, critically ill patients and demonstrated that an increase in SV > 12.5% correlated with fluid responsiveness.[^27]

Historically, the passive leg raise has been applied successfully in a variety of clinical scenarios, although the application of echocardiography with the passive leg raise is rather recent. One difficulty with transthoracic echocardiographic assessment of stroke volume or cardiac output in the passive leg raise is the possibility to alter the angle of insonation during the maneuver. If the heart experiences translational movement during the maneuver, the measured change in SV could be erroneous. Similarly, if the clinician does not ensure that the probe angulation and location are exactly the same before and after the maneuver, the measurements could be erroneous. The passive leg raise also may not be useful in patients with abdominal compartment syndrome or pregnancy.

See the One Minute Ultrasound for a demonstration of the passive leg raise here (#Movie 12.12).

Movie 12.12 One Minute Ultrasound Passive Leg Raise Demo

12.10 Recommendations

Highlights

In critical patients where too much fluid can kill, ultrasound can help you decide if fluid is the answer.

Ultrasound is a useful tool for the clinician faced with the dilemma of whether or not he or she should administer a fluid challenge. In patients who are receiving passive ventilation from mechanical ventilation, the degree of stroke volume variation and the degree of IVC collapse correlate with likelihood to respond to fluid. In spontaneously breathing patients, an increase in SV following a passive leg raise is a good method to predict fluid response. Ultrasound may also be useful in detecting elevated left atrial pressure and pulmonary edema, a frequent complication of fluid administration.

12.11 References


  1. Wiedemann HP, Wheeler AP, Bernard GR, et al. Comparison of two fluid-management strategies in acute lung injury. N Engl J Med 2006; 354:2564-2575. ↩︎

  2. Durairaj L, Schmidt GA. Fluid therapy in resuscitated sepsis: less is more. Chest 2008; 133:252-263. ↩︎

  3. Maitland K, Kiguli S, Opoka RO, et al. Mortality after fluid bolus in African children with severe infection. N Engl J Med 2011; 364:2483-2495. ↩︎

  4. Starling EH, Visscher MB. The regulation of the energy output of the heart. J Physiol 1927; 62:243-261 ↩︎

  5. Editorial: Starling's law survives. Lancet 1974; 2:818. ↩︎

  6. Osman D, Ridel C, Ray P, et al. Cardiac filling pressures are not appropriate to predict hemodynamic response to volume challenge. Crit Care Med 2007; 35:64-68. ↩︎

  7. Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest 2008; 134:172-178. ↩︎

  8. Shippy CR, Appel PL, Shoemaker WC. Reliability of clinical monitoring to assess blood volume in critically ill patients. Crit Care Med 1984; 12:107-112. ↩︎

  9. Marik PE, Monnet X, Teboul JL. Hemodynamic parameters to guide fluid therapy. Ann Intensive Care 2011; 1:1. ↩︎

  10. Grinnan DC, Truwit JD. Clinical review: respiratory mechanics in spontaneous and assisted ventilation. Crit Care 2005; 9:472-484. ↩︎

  11. Putensen C, Muders T, Varelmann D, et al. The impact of spontaneous breathing during mechanical ventilation. Curr Opin Crit Care 2006; 12:13-18. ↩︎

  12. Brennan JM, Blair JE, Goonewardena S, et al. Reappraisal of the use of inferior vena cava for estimating right atrial pressure. J Am Soc Echocardiogr 2007; 20:857-861. ↩︎

  13. Jue J, Chung W, Schiller NB. Does inferior vena cava size predict right atrial pressures in patients receiving mechanical ventilation? J Am Soc Echocardiogr 1992; 5:613-619. ↩︎

  14. Barbier C, Loubieres Y, Schmit C, et al. Respiratory changes in inferior vena cava diameter are helpful in predicting fluid responsiveness in ventilated septic patients. Intensive Care Med 2004; 30:1740-1746 ↩︎